Introduction
The monitor has many functions that you can control to monitor adults, children and newborns. It is powered by the published 100V – 240V ~ 50Hz/60Hz. Users can choose different configurations according to their requirements. Data and waveform are displayed while on the 5-inch TFT LCD monitor. It can be connected to the Central Monitoring System via a wireless network creating a network monitoring system.
Can monitor parameters such as ECG, RESP, SpO2, PR, NIBP, TEMP and CO2 (optional). It integrates the parameter measurement module and the display in one device, located in a portable monitor.
Function
Parameters: ECG, RESP, SpO2, PR, NIBP, TEMP, CO2 (optional)
1) ECG waveform of heart rate (HR).
2)RESP Respiratory Rate (RR) Respiration waveform
3) SpO2 Oxygen saturation (SpO2) Plethysmogram (PLETH)
waveform, pulse rate (PR), bar graph
4) NIBP systolic diseases (SYS), diastolic diseases (DIA), additional basic diseases (MEAN)
5)TEMP
6)CO2 (optional) EtCO2, INSCO2, AWRR
It has functions such as audible and functional alarm, provides and outputs trend data, NIBP measurement, alarm event marking and drug calculation.
Features
1) 5 inch TFT LCD display, multi-language interface (simplified Chinese, English, Russian, Kazakh, Slovak, Czech, Spanish, etc.).
2) Fanless design, quiet, power supply and clean, which reduces the possibility of cross-services.
3) With standard interface, OxyCRG, character interface and wire interface, convenient to solve.
4) Optionally, you can run waveform, parameter color and location.
5)All operations can be completed through the mechanism and touch screen.
6) Use a digital version of SpO2 that combines motion and light disturbances, and the device can be used in perfusion conditions.
7) NIBP measurement mode: Manual / AUTO / STAT, provide NIBP data for 2400 groups.
8) Connect to the central monitoring system via Wi-Fi.
9) Built-in large capacity battery, meets the requirements of final work. AC/DC ensures uninterrupted monitoring.
10) anti-high-frequency surgical unit, intended for defibrillation (special ones are necessary).
Performance
▲ECG
Routing mode: 3-wire: I, II, III
5-lead: I, II, III, aVR, aVL, aVF, V
Wave shape: 3 wires, 1 channel
5-wire, 2-channel
Gain: 2.5 mm/mV, 5.0 mm/mV, 10 mm/mV, 20 mm/mV
Scanning speed: 12.5 mm/s, 25 mm/s, 50 mm/s
HR
Measuring and alarm range:
To: rose 15 ~ 300 beats at the bottom
Children/Newborns: 15 ~ 350 beats per moment
Accuracy: ± 1% or ± 1 stroke per second, whichever is greater
Resolution: 1 bpm
▲ANSWER
Method: RF impedance (RA-LL).
Breathing rate:
Measuring and alarm range: 0 ~ 150 rpm /min
Resolution: 1 image /min
Accuracy:
7 rpm ~ 150 rpm, ± 2 rpm or ± 2%, whichever is greater;
0 rpm ~ 6 rpm, undefined
Apnea alarm: 10 ~ 40 s
Scanning speed: 12.5 mm/s, 25 mm/s
▲NIBP
Measurement mode: Oscillometer
Operating mode: Manual / AUTO / STAT
Measurement between measurements AUTO: 1 / 2 /3 / 4 / 5/10 / 15 / 30 / 60 / 90 / 120 / 180 / 240 / 480 minutes
STAT measurement period: 5 minutes
Measurement and alarm range: 0 ~ 300 mmHg
Resolution: 1 mmHg
Cuff accuracy: ± 3 mmHg
Measurement accuracy:
Maximum mean deviation: ± 5 mmHg
Maximum standard deviation: 8 mmHg
Upload protection:
Adult 297±3 mmHg
Children: 260±3 mmHg
Newborn: 147±3 mmHg
▲SpO2
Measurement and alarm range: 0 ~ 100%
Resolution: 1%
Accuracy: 70% ~ 100%: ± 2%;
0% ~ 69%: undefined
▲PR
Measurement and alarm range: 25 ~ 250 bpm
Accuracy: ± 2 beats per or ± 2%, whichever is greater
▲TEMP
Measurement and alarm range: 0 ~ 50℃
Resolution: 0.1℃
Accuracy: ± 0.1 ℃
▲Power supply: 100V-240V ~ 50Hz/60Hz
▲Safety class: Class II
Accessories
1) SpO2 probe on an adult's fingertip
1) Adult NIBP cuff
2)NIBP extension tube
2) ECG lead cable
3) ECG electrode
4) TEMP. Polls
3) Power cord
4) User manual
Features
Dimensions: 168 mm (L) × 56 mm (W) × 98 mm (H)
Weight: <1.0 kg (no storage required)